Comparative Study on the Results with Different Biochemical Detecting Systems

HE Qian,ZHAO Li-hua,WANG Xiang-ling,LIU Yu-jing,YIN Bin-xia
DOI: https://doi.org/10.3969/j.issn.1671-7414.2007.06.014
2007-01-01
Abstract:Objective To explore the comparability of results of different biochemical detecting systems.Methods According to EP9-A file of NCCLS,two biochemical detecting systems,Roche 7600 and HITACHI 7600,were used to detect serum biochemical results to obtain the correlation coefficient and the linear equation,respectively.Then the coefficient and equation were used to evaluate the system bias between laboratory method and comparison method.Results The outcome bias of ALB could be accepted partly between two systems,the biases of TBIL and DBIL were not acceptable.Conclusion The comparability of results in 2 detecting systems accorded to evaluation of clinical acceptability.