Comparison of different immunochemistry system based on National Committee for Clinical Laboratory Standards EP9-A2 document

TANG Jiangtao,WANG Lanlan,LI Lixin,CHEN Jie,CAI Bei
DOI: https://doi.org/10.3969/j.issn.1673-8640.2006.06.008
IF: 1.865
2006-01-01
Laboratory Medicine
Abstract:Objective To evaluate the difference between two IMMAGE immunochemistry systems in our laboratory and ensure the concordance of results in two IMMAGE. Methods We did this comparison in accordance to National Committee for Clinical Laboratory Standards(NCCLS) EP9-A2(method comparison bias estimation using patient sample) and the comparison contains 9 test items developed routinely in our laboratory. Results The correlation coefficient of 9 test items between two IMMAGE was higher than (0.975) and true bias at concentration X_C(Bc) was smaller than acceptable error(Ea).Conclusions The difference between two IMMAGE is in the clinical allowable range,our laboratory can provide the acute report for clinic.