Comparative analysis of α-amylase levels determined between eight clinical detection systems and reference method

HAN Li-qiao,ZHUANG Jun-hua,HUANG Xian-zhang,WANG Jian-bing,ZHENG Song-bai
2012-01-01
Abstract:Objective To evaluate the comparability of α-amylase(Amy) levels determined by different clinical detection systems and reference method.Methods The Amy levels in the sera of patients,domestic and international inter-laboratory comparison samples,and commercial calibration and quality control materials were determined by 8 clinical detection systems and reference method,simultaneously,and compared according to EP9-A2 protocol of American Clinical and Laboratory Standards Institute(CLSI).Then,the clinical acceptability of every detection system was evaluated according to the standard of the half of total error allowed by National Center for Clinical Laboratories.Results The correlation equations for serum samples between reference method(X axis) and clinical detection systems(Y axis 1-8) were Y1=0.934 7X-2.375 0,Y2=0.801 6X+13.901 4,Y3=1.139 7X+13.979 5,Y4 =0.829 3X+0.702 4,Y5=0.830 6X+5.826 9,Y6=0.791 0X+12.195 8,Y7=0.728 4X-0.826 5 and Y8=0.721 9X+44.318 0,respectively.Among them,the results of Y1,Y2,Y5 and Y6 detection systems could be clinically accepted,but the results of Y4 and Y7 systems must be calibrated with reference method for clinical acceptance.Conclusion Calibration for clinical detection system with reference method may increase the comparability of Amy levels detected with different detection systems.The establishment of the reference system should be the key to ensure the comparability of test results from different clinical detection systems.