Comparabilily Study on Results of TP and ALB with Different Detecting Systems

ZOU De-xue,XIA Yong,CHEN Yan-e,WU Xia-feng,WU Zhi-cheng
DOI: https://doi.org/10.3969/j.issn.1671-7414.2008.04.024
2008-01-01
Abstract:Objective The comparability of serum total protein and albumin results of four different biochemical detecting systems were studied through bias estimation and method comparison.Methods Following NCCLS document EP9-A2,detecting system including vitros 950 biochemical analyzer as comparison method,the rest three different biochemical detecting system as laboratory method and were used respectively to TP and ALB,then to obtain the correlation coefficient and the linear equation.Then the coefficient and the equation were used to evaluate the system bias between laboratory method and comparison method.Judged the comparability of different investigate systems according to the half of CLIA'88 standard.Results The correlation between systems was good(r>0.975);the results of TP,ALB of systems 1,2 and 3 were acceptable to the comparison method.Conclusion If the same test are determined by more than two systems,it is necessary to do method comparison in order to insure the comparability,accurate and consistent.