Study on the Traceability and Comparability of a Self-Developed Biochemical Testing System

ZHANG Xiuming,LI Weixuan,LAN Kai,ZHENG Songbai,CHEN Binhong,SUN Lei,LIN Lianying,LIANG Weixiong
DOI: https://doi.org/10.3969/j.issn.1673-8640.2007.03.024
IF: 1.865
2007-01-01
Laboratory Medicine
Abstract:Objective To establish a method to achieve the traceability and comparability of a self-developed biochemical testing system. Methods The fresh serum of patients was adopted as temporary calibrator and assayed in objective testing system before it was applied to calibrate self-developed testing system.After successful calibration verification,daily calibrators were determined in self-developed system and attained a new calibration value.Then this value of calibrator was used to calibrate self-developed testing systems again.After effective verification,we evaluated the bias between self-developed system and objective system by comparative test,and decided that the bias was acceptable or not. Results The bias of six items with no comparability between self-developed systems Ⅰ and objective system,and the bias of five items with no comparability between self-developed system Ⅱ and objective system,after calibration and verification,all the bias were lower than half of CLIA′88 allowable total error. Conclusions The traceability and comparability of self-developed biochemical testing system can be achieved by calibration and verification.
What problem does this paper attempt to address?