The Contrast and Bias Estimation on the Result of Routine Biochemistry Items in Different Biochemical Testing Systems

ZHANG Xiuming,LI Weixuan,ZHENG Songbai,CHEN Binhong,CAO Yongjian,SUN Lei,LIN Lianying,KE Peifeng,MA Ji
DOI: https://doi.org/10.3969/j.issn.1673-8640.2007.02.023
IF: 1.865
2007-01-01
Laboratory Medicine
Abstract:Objective To study the results comparability of different biochemical testing systems by bias estimation and method comparison. Methods According to National Committee for Clinical Laboratory Standards(NCCLS) document EP9-A2,17 biochemistry items in different self-built biochemical testing system and aim system were tested respectively to obtain the correlation coefficient and the linear equation.Then the system bias between self-built testing system and aim system could be evaluated by obtained equation.The comparability of different investigated system was judged according to the half of clinical laboratory improvement amendment 88(CLIA′88) allowed total error. Results The results of total bilirubin(TBil),direct bilirubin(DBil),γ-glutamy itransferase(GGT),triglycerides(TG),total calcium(TCa~(2+)) and inorganic phosphate(IP) tested in system 1 and aim system showed no comparability.The results of albumin(Alb),TBil,DBil,TG and IP tested in system 2 and aim system showed no comparability.There was comparability between system 3 and aim system in all items. Conclusions Routine biochemistry results had different bias between self-built testing system and Roche Modular PPI system.If the same item is tested in more than two systems,it is necessary to judge clinical acceptability by method comparison and bias estimation so as to insure the comparability.