Accuracy evaluation of Afinion AS100 hemoglobin A1c analyzer in point-to-care testing

LI Jin,DENG Hong-rong,YANG Dai-zhi,YAN Jin-hua,SHI Yi-ming,XING Xiao-huan,WENG Jian-ping
DOI: https://doi.org/10.3969/j.issn.1001-9448.2012.08.031
2012-01-01
Abstract:Objective To evaluate the accuracy of a new kind of point-of-care Hemoglobin A1c(HbA1C) testing analyzer.Methods Seventy diabetic patients and 12 health volunteers were enrolled.HbA1C was assessed with the same Afinion AS100 Analyzer(point-of-care testing,boronate affinity method) and 2 Bio-Rad D10 Analyzers(ion-exchange high performance liquid chromatography method).Moreover,the HbA1C was re-assessed with the same analyzers after whole blood specimens were refrigerated in 2~8℃ for 7 days.Results There was no significant difference in results of HbA1C among 3 analyzers(P>0.05),neither after refrigeration(P>0.05).Conclusion Afinion AS Analyzer provides similar accuracy and reliability with commonly used analyzer in clinical laboratory.
What problem does this paper attempt to address?