Evaluation of the Accuracy of the Fructosamine Level Tested by Cobas INTEGRA 400 Analyzer

LIANG Hong,CAI Xiao-ling,JI Li-nong
DOI: https://doi.org/10.3969/j.issn.1674-1633.2006.10.016
2006-01-01
Abstract:Objective To evaluate the accuracy of the fructosamine level tested by cobas INTEGRA 400 analyzer.Methods The level of fructosamine of 105 cases of type 2 diabetes patients was tested by cobas INTEGRA 400 analyzer and GlyPro analyzer.Results The correlation analysis showed that the level of fructosamine tested by both the two analyzers were significantly correlated(r=0.925,p0.05). And the level of fructosamine tested by cobas INTEGRA 400 analyzer and GlyPro analyzer was also significantly correlated with hemoglobin A1c level respectively(r=0.816,0.790,p0.05).Conclusion The fructosamine level tested by cobas INTEGRA 400 analyzer was accurate.