Evaluation of the Medisafe glucometer' s performance

WANG Hai-ning,CHEN Li,HONG Tian-pei
DOI: https://doi.org/10.3969/j.issn.1006-6187.2009.03.025
2009-01-01
Abstract:Objective To evaluate the performance of strip-Medisafe glucometer(Glucometer).Methods Vein blood samples of 220 patients were tested by both Medisafe glucometer and automatic biochemistry analyzer, and the correlation between the testing results of two methods was analyzed. Precision and linearity of the glucometer were evaluated, and effects of hematocrit (HCT) variation on the glucometer's performance were investigated. Results The whole blood readings of the glucometer (Y) were well correlated with the plasma glucose levels of the biochemistry analyzer (X), with the linear equation of Y= 0.9731X (r~2=0.9851). Error analysis showed that the difference between two methods was within 0%~18%. The precision test indicated that the within-lot coefficient of variation (CV) of the strip-glucometer was 2.3%~3.6%, and the lot-to-lot and meter-to-meter CVs were 2.8~3.5% and 3.1%~3.9% respectively. When the HCT level of blood samples ranged from 20% to 60%, the reading difference between the glucometer and the analyzer were less than 18.4%. Conclusions Medisafe glucometer holds nice accuracy and precision, wide range of linearity, and good ability of anti-jamming. It is suitable for the glucose testing both in hospital sickroom and at home by the diabetic patients.