Performance verification of affinity chromatography HPLC in HbA1c detection

Yu CHEN,Juan DU,Ying-ying HU,Zhi-hong QI,Hui ZHANG,Wei CUI
2014-01-01
Abstract:Objective To verify the analytical performance of affinity chromatography HPLC in haemoglobin A1c(HbA1c) detection. Methods The accuracy,imprecision,detection range,contamination carryover rate,methodology comparison and interference of affinity chromatography HPLC were verified referring to the CLSI evaluation protocols EP5,EP6 and EP9.Results Accuracy test results were all within the range of quality controls,and close to the target values. Imprecision results at three different levels were 0. 82%,0. 72% and 0. 79%,respectively,which met the requirement(less than 1%). The HbA1 c showed good linear correlation in the concentration range of 5. 2% ~ 18. 8%( y = 1. 003x-0. 125,r = 0. 995). The carryover value was 0. 12 when HbA1 c concentration in the range of 4. 9% ~ 15%,the cross contamination between samples can be ignored. Manual dilution evaluation test results showed that the maximum deviation between whole blood and dilution mode was0. 1. Premier Hb9210 and VariantⅡ were in good correlation,with the linear regression equation y = 1. 0001x-0. 003( r =0. 998). Paired t-test data of the two groups(n = 40) showed there was no significant statistical difference(t = 0,P = 1. 43).In interference test,the results of affinity chromatography HPLC method(Premier Hb9210) had good correlation with the blood glucose(group HbJ,r = 0. 927; group HbF,r = 0. 827),and the results of VariantⅡ had poor correlation with the blood glucose(group HbJ,r = 0. 024; group HbF,r = 0). Conclusion The HbA1 c detection by affinity chromatography HPLC method is in line with the requirements of CLSI,applicable to routine clinical testing.
What problem does this paper attempt to address?