The Application of IFCC Reference Methods in Methodological Evaluation of Glycated Hemoglobin Measurement Systems

Zhang Tianjiao,Pu Yungang,Zhou Huijuan,Ma Rong,Zhang Jiangtao,Wang Daguang,Li Chuanbao,Li Miao,Wang Yufei,Zhang Chuanbao
DOI: https://doi.org/10.3760/cma.j.issn.1009-9158.2018.11.007
2018-01-01
Abstract:Objective To perform a methodological evaluation study of 4 HPLC based systems and a capillary electrophoresis based system , with the International Federation of Clinical Chemistry and Laboratory Medicine ( IFCC) glycated hemoglobin reference method as a comparative method .Methods 40 hemolysis samples of variety concentrations were prepared .The samples were measured by IFCC reference method and 5 glycated hemoglobin testing systems , respectively and trueness verification was performed according to the Clinical &Laboratory Standards Institute (CLSI) guideline EP9-A3.Whole blood samples were used to test the systems'precision, linearity and analytical interferences .Results The average CV of IFCC reference method results of 40 hemolysis samples was 1.4%( range from 0.2%to 2.5%).No outlier was found in the results of the 5 testing system.The slopes ranged from 0.9902 to 1.0267, and intercepts from -0.1526 mmol/mol to +0.1512 mmol/mol, squared correlation coefficient from 0.9962-0.9971, biases at two medical decision level were less than 0.3%HbA1c.Within-laboratory precisions were less than 2% (NGSP unit).Bias between all test results and predicted results were less than 5%.High concentration of glucose showed certain interference to glycated Hemoglobin tests but had little influence in clinical practice.Conclusions The results of the 5 testing system are comparable to the IFCC reference method , the results of precision and linearity evaluation meet the requirements of related guidelines .
What problem does this paper attempt to address?