The Comparative Analysis for the Detection of Human Serum Total β-chorionic Gonadotropin between Beckman Coulter Access DXI800 and Cobas e411

Xue-jun QIN,Min-zhi AN,Yang-yu LI,Qi-guan MAO,Qing LIN
DOI: https://doi.org/10.16662/j.cnki.1674-0742.2016.23.033
2016-01-01
Abstract:Objective To Compare two different chemiluminescene assays (Beckman Coulter Access DXI800 and Cobas e411) for the detection of human serum total β-chorionic gonadotropin (β-HCG)and discuss they are comparable between the two different ECL assays, in order to provide evidence for mutual accredit between different assay systems. Methods Biorad low and median value QC were used for Beckman Coulter Access DXI800 and Roche original low and median value QC were used for Cobas e411 daily, guaranteed all the results were in control. On this basis, 40 different serum samples were compared of β-HCG by Beckman Coulter Access DXI800 (candidate method) and Cobas e411 (reference method), all data were tested in 2 hours. Results The different QC daily CV and total CV for β-HCG tested by two analyzers were less than 15%of the day CV allows the scope of . Data of 40 different concentrations of β-HCG in two test systems were paired by t test, showed no significant difference (P=0.563>0.05). Correlation analysis showed the coefficient r2 was 0.9936>0.975, in line with the requirements of the EP9-A2 file of the American clinical chemistry laboratory standardization committee, which met the requirements of the experiment. Conclusion As results, two kinds of detection system for β-HCG test results had good accuracy and correlation, avoided the trouble of clinical diagnosis caused by the differences as two results were acceptable.