Comparison study and analysis of detecting HBsAb using ECLIA and ELISA

ZHAO Xiu-ying,CHEN Jun-mei,WANG Wen-jing
2011-01-01
Abstract:Objective To evaluate the difference between electrochemi-luminescence immuno-assay(ECLIA) and enzyme-link immunosorbent assay(ELISA) in detecting Hepatitis B surface antibody(HBsAb).Methods Based on ROCHE E601 ECLIA method,65 serum samples of both HBsAg and HBsAb positive,63 serum samples of HBsAg negative and HBsAb positive were selected.Two ELISA reagents(A and B) were used to detect HBsAb again.Results When HBsAg was positive,both ELISA reagents missed the diagnosis in HBsAb,with the relative missing rate at 84.62% and 78.46%.There was no obviously improvement when repeated the experiment with the diluted serum.Whereas,when the HBsAg was negative,the relative missing rate for reagents A and B in detecting HBsAb were 4.76% and 7.94%.Conclusions There exists an obviously influence from HBsAg to serum HBsAb assay using ELISA method.It is necessary to improve the efficiency of HBsAb kits,to reduce the interference from HBsAg.It is recommended to use a luminescence reagent to method to detect the HBsAb,especially for those patients being diagnosed with Hepatitis B infection and people who need injection of Hepatitis B vaccine.