Evaluation of different methods in determination of low level HBsAg]

Chun-rong Fei,Ai-qing Ye,Jun Zhang
DOI: https://doi.org/10.3785/j.Issn.1008-9292.2011.04.016
2011-01-01
Abstract:To evaluate chemiluminescent immunoassay (CLIA), electrochemiluminescent immunoassay (ECLIA) and ELISA in determination of low level HBsAg.According to the standard of CLIA Architect i2000, 70 samples were divided into three groups by HBsAg concentration : <1 ng/ml, 1-5 ng/ml and >4 ng/ml. The samples were also determined by ECLIA MODULAR<E>170 and ELISA, and the results were compared with those measured by CLIA Architect i2000.The concordance rates of ECLIA MODULAR<E>170 with Architect i2000 was 79.2% for <1 ng/ml group, 100% for 1-5 ng/ml group, 100% for >4 ng/ml group. And the concordance rates of ELISA with Architect i2000 was 0 % for <1 ng/ml group, 45.5% for 1-5 ng/ml group and 100 % for >4 ng/ml group.For determination of low level HBsAg,CLIA Architect i2000 and ECLIA MODULAR<E>170 have high credibility. ELISA is also credible when HBsAg >5 ng/ml, but not for low level HBsAg, especially for HBsAg <1 ng/ml.
What problem does this paper attempt to address?