Determination of Ni, Cu and Hg by flow injection preconcentration coupled with HPLC
Akbar Jahubar Ali,Hong Shen,Guangming Xu,Xuefeng Yin
DOI: https://doi.org/10.3321/j.issn:0251-0790.1998.11.032
1998-01-01
Abstract:A home designed pressure tight preconcentration column combined with the HPLC six-port valve was used to interface the FI sorbent extraction preconcentration system and HPLC, Based on this system, a method for determination of trace amount of Ni, Cu and Hg in water samples after on-line derivatization and preconcentration as their diethyldithiocarbamate complexes has been developed, The sensitivity and selectivity can be further improved by using time programming for absorption measurement. The detection limit 0.16 mu g/L, 0.41 mu g/L and 1.1 mu g/L, and the relativity standard deviation, 3.6%, 4.5% and 3.9%, were found for Ni(II), Cu(II) and Hg(II), respectively.