Flow Injection On-line Photochemically Spectrofluorimetric Determination of Folic Acid in Pharmaceutical Preparations

CB Huang,HW Chen,QH He
DOI: https://doi.org/10.3321/j.issn:0253-3820.2003.02.025
2003-01-01
Abstract:Upon exposure to ultraviolet-irradiation folic acid was transformed to a strongly fluorescent compound with a maximum excitation wavelength of 274 nm and a maximum emission wavelength of 466 nm. Investigation into them medium for the photochemical reaction revealed that the most intensive photochemical fluorescence was induced in Na2CO3-NaHCO3 buffer. Based on these observations, a flow injection on-line photochemical spectrofluorimetry was developed. A sample band, carried by a water stream, was merged with a pH 9.5 Na2CO3 - NaHCO3 buffer stream. The mixed solution was driven to pass a knotted polytetrafluoroethylene photochemical reactor that was freely coiled around a 6-W low-pressure mercury lamp. The fluorescent compound produced in the reactor Was directly delivered to a flow-through-cell where the fluorescence intensity was measured. Under the optimized conditions a detection limit of 0.16 mug/L was achieved at the sampling rate of 80/h. Eleven determinations of standard solutions containing both 10 and 100 mug/L folic acid gave RSDs of 1.0% and 0.2%, respectively. Linear dynamic range was observed in the range of 1 similar to 4000 mug/L. The developed approach was successfully applied to the determination of folic acid in tablets.
What problem does this paper attempt to address?