Determination of Iron( II ) and Total Iron Using Flow Injection On-line Preconcentration with FAAS

Guangming Xu,Yingxue Ye,Xuefeng Yin,Hong Shen,Akbar Jahubar Ali
2000-01-01
Abstract:A direct, rapid and sensitive method for the determination of Fe(II) and Fe(III) in water samples by flame absorption spectrometry coupled with flow injection on-line preconcentration with 1,10-phenanthroline (1,10-phen) as the complex agent was developed. Various parameters affecting the complex formation and its subsequent adsorption, such as pH of the mixture solution, 1,10-phen concentration, sample flow rate etc. were optimized. A 2.5 X 10(-3) mol/L 1,10-phen in 2.0% ethanol was on-line mixed with aqueous sample solution of iron in 0.1% (volume ratio) hydrochloric acid and flowed through the microcolumn for 30 s. Fe(II) can be extracted as Fe(II)-phen complex in the C-18 microcolumn and determined within 10 s in the nebulizer of FAAS after being eluted by ethanol. Ascorbic acid (0.1 mg/mL) was used for the reduction of Fe(III) to Fe(II). A good precision (RSD = 1.1%, n = 10), high enrichment factor (19) and sample throughput (80 h) with detection limit 3 mu g/L were obtained. The method has been successfully applied on the water sample for Fe(II) and Fe(III) determination with the recovery 94%-105%.
What problem does this paper attempt to address?