Determination of Pb and Cd in water sample by on-line ion-exchange preconcentration-FAAS

LI Dan,YU Xiaofeng,SHOU Miao-jun,YE Hua-jun,HAN Shuang-lai,ZHOU Jin-fan
DOI: https://doi.org/10.13822/j.cnki.hxsj.2013.02.012
2013-01-01
Abstract:Trace amount of Pb and Cd was determined using an on-line ion-exchange column preconcentration Flame atomic absorption spectroscopy system(IE-FAAS).The effects of experimental variables such as the pH of sample,the content of ion-exchange resin and coexisting ions on the recovery of Pb and Cd were disscused.The optimum operation conditions were as follows: the content of ion-exchange resin was 0.50 g,the pH of solution was 6.0,the eluent was 0.80 mol / L HNO3.The proposed method could be used for on-line separation and preconcentration of Pb and Cd in water sample.The sensitivity of IE-FAAS was 100 times higher than that of FAAS.Recoveries obtained by standard addition method were in the range from 90% to 102%.The results demonstrate that the proposed on-line IE-FAAS has the advantages of easy operation,low cost and sensitivity.It can be applied to in situ measurement of trace metals in water sample.