Determination of trace copper and nickel in environmental and biological samples by flow injection on-line microcolumn preconcentration flame AAS using acrylic acid-grafted polytetrafluoroethylene fiber for column packing

ZhaoHui Wang,Zhipeng Wang,Zhengpu Zhang,LiWen Liu,Xiuping Yan
2005-01-01
Atomic Spectroscopy
Abstract:A newly prepared acrylic acid-grafted PTFE fiber sorbent (see reference 20) was extended to flow injection (FI) on-line micro-column preconcentration - flame atomic absorption spectrometric determination of trace copper and nickel in environmental and biological samples. On-line preconcentration of trace analytes was achieved on the microcolumn packed with acrylic acid-grafted PTFE fibers, and the retained analytes were eluted with 1.0 mol L-1 HCl for on-line. FAAS determination. The optimum pH range of the sample solution for preconcentration of Cu(II) and Ni(II) was 3.6-5.6 and 4.4-6.4, respectively. The developed sorbent exhibited fairly fast kinetics for the adsorption of Cu(II) and Ni(II), permitting the use of high sample flow rates up to at least 10.8 ml, min(-1) for the FI on4ine microcolumn preconcentration system without loss of the retention efficiency. With a preconcentration time of 45 s at a sample loading flow rate of 10 mL min(-1), an enhancement factor of 48 for Cu and 23 and a detection limit (3sigma) for Ni, of 0.20 mug L-1 for Cu and 0.25 pg L-1 for Ni were achieved at a sample throughput of 55 h(-1). Precision for 11 replicate measurements of 25 mug L-1 Cu and 30 mug L-1 Ni was 1.2 and 1.6% (RSD), respectively. The method was successfully applied to the determination of trace copper and nickel in environmental and biological samples.