Iron Determination after On-Line Sorbent Preconcentration by Fi-Fass Using 1, 10-Phenanthroline As a Complexing Agent

Akbar Ali,Ye Ying-xue,Xu Guang-ming,Zhang Tao,Yin Xue-feng,Jin Xin
DOI: https://doi.org/10.1631/bf02841043
2000-01-01
Journal of Zhejiang University SCIENCE A
Abstract:1, 10-Phenanthroline (phen) was tested as a complexing agent for on-line preconcentration of iron on RP-C 18 material in a microcolumn with flow injection coupled with flame atomic absorption spectrometer (FI-FAAS). The on-line formed Fe(II)-phen complexes were adsorbed on the C 18 sorbent material. Various parameters affecting the on-line Fe(II)-phen complex formation and its subsequent adsorption in the microcolumn as well as its elution into the nebulizer of AAS were optimized. A 2.5×10 −3 mol/L phen in 2% through the microcolumn for 30 s. The adsorbed Fe(II)-phen complexes in the microcolumn were eluted with ethanol in 10 s into the nebulizer of AAS. Ascorbic acid (5×10 −4 mol/L) was added to the sample solution for the reduction of Fe(III) to Fe(II). A good precision (RSD=1.1%, n =10), high enrichment factor (19) and sample throughput (90 h −1 ) with detection limit (3σ) of 3 μg/L were obtained. The method was applied on standard reference materials (i. e. mussel and tomato leaves,) for iron determination and yielded results agreeing well with certified values.
What problem does this paper attempt to address?