A Sensitive Fluorescence Quenching Method For The Determination Of Iron(Ii) With 1,10-Phenanthroline

Guanghua Zhu,Huangxian Ju,Baofen Ye
DOI: https://doi.org/10.1002/cjoc.20030210317
IF: 5.56
2003-01-01
Chinese Journal of Chemistry
Abstract:A sensitive and selective fluorescence quenching method for the determination of Fe2+ with 1,10-phenandwoline was developed. The fluorescence intensity of 1, 10-phenanthroline at lambda(ex) of 266 nm and lambda(em) of 365 nm was constant in the range of pH 4.0 to 10.0 and decreased linearly upon addition of Fe2+ to its solution. This decrease was mainly due to a static quenching effect caused by the formation of a non-fluorescent complex of Fe2+ with 1, 10-phenanthroline. The total amount of iron was determined by using hydroxylamine hydrochloride to reduce Fe3+ to Fe2+. The linear range was from 5.0 x 10(-7) to 2.0 x 10(-5) mol/L with a detection limit of 2.4 x 10(-8) mol/L at 3sigma. The quenching constant of Fe2+ to 1,10-phenanthroline was calculated to be (5.70 +/- 0.05) x 10(4) L/mol at 25 degreesC. Effects of foreign ions on the determination of Fe2+ were investigated. The results of the new method for the determination of iron in tap water and natural water samples were in good agreement with those obtained by graphite atomic absorption spectrometry.
What problem does this paper attempt to address?