Analysis of Samples with High Dissolved Solids Content Using Flow-Injection Flame Atomic-Absorption Spectrometry

Z FANG,B WELZ,G SCHLEMMER
DOI: https://doi.org/10.1039/ja9890400091
1989-01-01
Journal of Analytical Atomic Spectrometry
Abstract:Flow injection (FI) sample introduction in flame atomic absorption spectrometry (AAS) was used to analyse samples containing high concentrations of dissolved solids. The Fl system was optimised with respect to carrier flow-rate, nebuliser uptake rate, sample volume and sampling frequency using 30%m/V NaCl solution with 1 mg l–1 Cu. Sensitivity and precision were comparable to those obtained using simple aqueous standards over a broad range of operating parameters. With a 6 ml min–1 carrier flow-rate, 10.5 ml min–1 uptake rate, 330-µl sample volume and 125 h–1 sampling frequency, the precisions (RSD) obtained for 1 mg l–1 of Cu in 30%m/V NaCl and in saturated lithium borate solutions and for 5 mg l–1 Cu in 100%m/V K2HPO4 solution were 2.0, 1.5 and 2.0%, respectively. Repetitive introduction of a saturated lithium borate solution for 465 injections resulted in no deterioration in performance. After 167 consecutive injections of 30%m/V NaCl solution there was a 10% decrease in the peak signal but no degradation in precision. Limiting factors of Fl-AAS and reasons for the tolerance of high dissolved solids in samples are discussed.