Expansion of Dynamic Range of Flame Atomic Absorption Spectrometry by an Efficient Flow Injection Dilution System Based on Dispersion of Microliter-Volume Samples

ZL FANG,B WELZ,M SPERLING
DOI: https://doi.org/10.1021/ac00061a009
IF: 7.4
1993-01-01
Analytical Chemistry
Abstract:An automatic on-line flow injection dilution system for flame atomic absorption spectrometry was developed based on the injection of low- and submicroliter sample volumes and their dispersion in a mixing coil. Precise control of volumes was possible using computer-controlled stepper motor-driven peristaltic pumps and small-bore Neoprene pump tubes. A more than 1000-fold dilution with a precision of better than 2% (rsd) (n = 15) was achieved using a sample volume of 0.7 muL and a mixing coil of 160-cm length and 1.3-mm i.d. Precision improved to 0.5-0.6% (rsd) when sampling times longer than 2 s were used. Sampling frequencies of 60-100/h, even for the highest dilutions, made the system highly compatible with routine application. A concentration of 7.2% Mg in an aluminum alloy was determined directly in the solution of 1 g of alloy in 100 mL.