Flame Atomic Absorption Determination of Lead in Biological Materials Using A Flow Injection On-Line Separation and Preconcentration System Based on Ion-Pair Sorbent Extraction

GH Tao,ZL Fang
1996-01-01
Atomic Spectroscopy
Abstract:A method was developed for the trace determination of lead by flame atomic absorption spectrometry with separation and pre-concentration by flow injection on-line ion-pair sorbent extraction. Lead was conveted into its iodoplumbate complex anion, thereby allowing it to be adsorbed on-line on an incorporated C-18 microcolumn as its ion-pair with tetrabutylammonium cation, and subsequently to be eluted by ethanol directly into the nebulizer burner system of the spectrometer. At a consumption of 7-mL sample volume, the procedure resulted in an enhancement factor of 36 and yielded a detection limit of 3 mu g/L (3 sigma) at a sampling frequency of 48/h. The precision (relative Standard deviation, n = 11) was 2.0% at the 300-mu g/L level and 3.1% at the 50-mu g/L, level. The method demonstrated high tolerance to interferences and results obtained for standard reference biological materials were in good agreement with the certified values.
What problem does this paper attempt to address?