Combination of Flow-Injection Techniques with Atomic Spectrometry in Agricultural and Environmental Analysis

ZL FANG,SK XU,X WANG,SC ZHANG
DOI: https://doi.org/10.1016/s0003-2670(00)84477-3
IF: 6.911
1986-01-01
Analytica Chimica Acta
Abstract:Combinations of flow-injection techniques with flame atomic absorption spectrometry (a.a.s.) and inductively-coupled plasma/atomic emission spectrometry (i.c.p./a.e.s.) are reviewed in the general context of agricultural and environmental analysis. The flow-injection systems are valuable for sample introduction; appropriate dispersion control allows the analysis of solutions containing as much as 40% (w/v) urea or phosphate in fertilizers. A study on the determination of cadmium in soil extracts by on-line ion-exchange preconcentration and flame a.a.s. detection is described. The interpolative standard-addition method with i.c.p./a.e.s. detection is outlined. Improvements in the determinations of selenium in environmental samples by hydride-generation a.a.s. and of mercury by cold-vapour a.a.s. are reported.
What problem does this paper attempt to address?