Determination of Trace Amounts of Copper by Flow Injection Vapor Generation AAS

HY Zhou,SK Xu,ZL Fang
DOI: https://doi.org/10.3321/j.issn:1000-0593.2000.04.018
2000-01-01
Abstract:A method for the determination of copper by FI-VGAAS (flow injection vapor generation atomic absorption spectrometry) was developed. The sample solution containing trace amounts of o-phenanthroline and 0.1 mol.L-1 HNO3 was carried out by carrier of 0.1 mol.L-1 HNO3 and merged with sodium tetrahydroborate solution to produce the vapor species of copper, which got separation in gas-liquid separator and was detected in a quartz tube atomizer at 1000 degrees C. A detection limit of 1.8 micrograms.L-1 (3 sigma) of copper was obtained with a sampling frequency of 144/h using a sample volume of 500 microL. The precision RSD was 2.6% (n = 11) at the 100 micrograms.L-1 copper. The method was applied to the analysis of human hair and rice standard reference samples, the results were in good agreement with certified values.
What problem does this paper attempt to address?