Measurement of the Wall Correction Factor of a Spherical-cylindrical Cavity Ionization Chamber

Jia-cheng HU,Xiao-yuan YANG,Yuan-di YANG,Hui ZHANG,Tie-shuan FAN,Ji-ning SUN,Fei-zeng HUANG,Jian-chang LIU,Wei-guo YE
2007-01-01
Abstract:An experimental method is used to determine the wall correction factor K wall of the spherical-cylindrical standard cavity ionization chamber, its volume is 10 cm 3, in 60Co gamma-ray air kerma measurements in the National Institute of Metrology. The factor K wall is determined by both the methods of traditional geometry thickness extrapolation and equivalent mean thickness extrapolation. The wall correction factor from extrapolation experiments for the spherical-cylindrical chambers is compared with the Monte Carlo (MC) calculations. It is concluded that the wall correction factor determined by traditional geometry thickness extrapolation method is less than 0.48% and 0.45% respectively as compared with that by the MC calculation and equivalent mean thickness. The results applying the equivalent mean thickness extrapolation and that of the MC calculations is agreed within 0.02%.