Assessment of four dose calculation algorithms using IAEA-TECDOC-1583 with medium dependency correction factor (K med ) application

Aram Rostami,Abdul Sattar Khalid,Hamed Ghafari,Satheesh Prasad Paloor,Bevan Orville Peltier,Rabih Hammoud,Shihab Abdelrahman
DOI: https://doi.org/10.1016/j.ejmp.2024.103390
IF: 3.119
2024-06-04
Physica Medica
Abstract:Highlights • The study highlights ion chamber limitations of dose measurement in lung and bone. • Medium correction factor improves accuracy in lung and bone dose measurements. • Study emphasizes the need for algorithm-specific detector corrections factor. Purpose This study discusses the measurement of dose in clinical commissioning tests described in IAEA-TECDOC-1583. It explores the application of Monte Carlo (MC) modelled medium dependency correction factors (K med ) for accurate dose measurement in bone and lung materials using the CIRS phantom. Methods BEAMnrc codes simulate radiation sources and model radiation transport for 6 MV and 15 MV photon beams. CT images of the CIRS phantom are converted to an MC compatible phantom. The PTW 30013 farmer chamber measures doses within modeled CIRS phantom. K med are determined by averaging values from four central voxels within the sensitive volume of the farmer chamber. K med is calculated for D m.m and D w.w algorithm types in bone and lung media for both photon beams. Results Average modelled correction factors for D m.m calculations using the farmer chamber are 0.976 (±0.1 %) for 6 MV and 0.979 (±0.1 %) for 15 MV in bone media. Correspondingly, correction factors for D w.w calculations are 0.99 (±0.3 %) and 0.992 (±0.4 %), respectively. For lung media, average correction factors for D m.m calculations are 1.02 (±0.3 %) for 6 MV and 1.022 (±0.4 %) for 15 MV. Correspondingly, correction factors for D w.w calculations are 1.01 (±0.3 %) and 1.012 (±0.2 %), respectively. Conclusions This study highlights the significant impact of applying Kmed on dose differences between measurement and calculation during the dose audit process.
radiology, nuclear medicine & medical imaging
What problem does this paper attempt to address?