Monte-Carlo Calculation and Measurement of Wall Correction Factor for Spherical Graphite Ionization Chamber

Liu Jianchang,Ji-ning Sun,Ye Weiguo,Huang Feizeng,Fan Tieshuan,Yang Xiaoyuan,Bin Guo,Yang Yuandi,Hu Jiacheng
IF: 1.279
2006-01-01
Nuclear Instruments and Methods in Physics Research Section B Beam Interactions with Materials and Atoms
Abstract:The factor K_(wall) to correct for photon attenuation and scatter in the wall of the 30 cm~3 and 10 cm~3 spherical chamhers at National Institute of Metrology for ~(60)Co air kerma measurement is determined by Monte-Carlo calculation and experimental meas- urement.The calculation using the EGSnrc software system shows that the calculated wall correction factors differ by up to 1.0% from those obtained hy linear extrapolations of the measured relative ionization current to zero wall thickness for the measured cham- bers.The relative ionization current and wall correction factors for the 30 cm~3 and 50 cm~3 spherical graphite chambers at NIST are calculated and compared with the meas- ured results.The Monte-Carlo code can reproduce the experimental relative ionization current within an accuracy of±0.12%.This work confirms that for spherical standard chambers the ~(60)Co air-kerma reference values should be obtained by a new K_(wall) deter- mined by Monte-Carlo calculations.