Effect of Probe-Gas Type and Pressure and Calibration Method on Leakage Rate Calibration of Deuterium

Wei Wang,Boyang Zhong,Xiaoqiu Ye,Fangfang Li,Chang' Chen,Tao Gao
DOI: https://doi.org/10.13922/j.cnki.cjovst.2017.12.02
2017-01-01
Abstract:The effect of the gas-type,pressure and calibration method on the helium and deuterium leakage-rate calibration with the commercial leak (2.3 ×10 -6 Pa· m3/s,He,23℃) was investigated in He-leak detection and constant volume methods.The results show that the probe-gas type,inlet pressure and calibration technique all have a major impact.For example,the leakage-rate increased with an increase of the inlet pressure,proportional to the quadratic difference of the inlet and outlet pressures.At the same inlet pressure,the leakage-rate of D2 was high-er than that of He;and an increase of the inlet pressure widened the difference.We found that the probe-gas of D2 resulted in a larger leakage-rate measured in He-leak detection technique than that evaluated in constant volume method,though the probe-gas of He gave the same leakage-rate.The correlation between the results rapidly meas-ured with the two probe gases was derived.