Research on Improving the System Sensitivity for Hot Helium Leak Test

Zhen Leng,Jiming Chen,Kun Wang,Min Xu,Xin Chen,Wei Luo,Renjun Fang,Pinghuai Wang
DOI: https://doi.org/10.1016/j.fusengdes.2023.113422
IF: 1.905
2023-01-01
Fusion Engineering and Design
Abstract:Hot Helium Leak Test (HHLT) can realize more reliable leak tightness assessment than the conventional cold helium leak testing for components that run at elevated temperatures. To ensure successful operation of the International Thermonuclear Experimental Reactor (ITER), the in-vessel components are required to ensure high vacuum sealing performance by HHLT. The biggest challenge for the HHLT is that the background leak rate and sensitivity of the test system can hardly meet the requirements of high accuracy at high temperature. Through years of research, it is found the main factor affecting the rate and the sensitivity of HHLT facility is the hydrogen released from the vacuum-facing materials at high temperature. By adopting a series of measures, the back-ground leakage rate was successfully reduced by one and a half order of magnitude to 1.17x10-10 Pa center dot m3/s at 253 degrees C, and the system sensitivity was improved by more than one order of magnitude to reach the order of 10-12 Pa center dot m3/s, even down to 9.85x10-13 Pa center dot m3/s at the temperature.