THE CONSTRUCTING OF THE HIGH PERFORMANCE ADIABATIC CALORIMETER

刘北平,沙伯海,谭志诚,孙立贤,颜君彪
DOI: https://doi.org/10.3969/j.issn.1672-6164.2003.01.014
2003-01-01
Abstract:A fully automated adiabatic caloumeter used for heat capacity measurements in the temperature range from 80 to 400K is constructed.All of the calorimetric procedures,including energy input,temperature measurement and adiabatic control,are perfoemed by the computer.The hardware of the calouimetric system consists of Agiednt 34970A Data Collection/Switch Unit and P4 computer.The reliability of the calorimetric system is verfied by measuring the heat capacities of the calorimetric standard material,α-Al 2O 3.The results indicate that the standard deviation of the data obtained by this instrument was within ±0.35% compared with those of the National Bureau of Standards(NBS).