Adsorptive microcalorimeter and its volume calibration

Jianyi Shen,Gufeng Wang
IF: 16.5
1998-01-01
Chinese Journal of Catalysis
Abstract:The principle and apparatus construction for microcalorimetric adsortpion are outlined. The key to the accuracy of measurement of differential adsorption heat is the dead volume calibration for cell section. Since the dosing and cell sections are usually held at different temperatures, a γ factor should be used to correct the effect of temperature difference, otherwise, a pseudo dead volume is obtained. By estimating the errors, it is found that both methods of using the γ factor and the pseudo volume give accurate results when temperature variations for the sections are negligible. However, when the temperature shift occurs for either dosing or cell section, the γ factor is more reliable to use than the pseudo volume in determining the adsorption amount. The γ factor can be obtained by measuring the dead volume of the cell section at the room and adsorption temperatures, respectively.