A Low-temperature Automated Adiabatic Calorimeter Heat Capacities of High-purity Graphite and Polystyrene

Zhi-Cheng Tan,Ji-Biao Zhang,Meng Shang-He
DOI: https://doi.org/10.1023/A:1010177331806
IF: 4.755
1999-01-01
Journal of Thermal Analysis and Calorimetry
Abstract:A computerized adiabatic calorimeter for heat capacity measurements in the temperature range 80–400 K has been constructed. The sample cell of the calorimeter, which is about 50 cm 3 in internal volume, is equipped with a platinum resistance thermometer and surrounded by an adiabatic shield and a guard shield. Two sets of 6-junction chromel-copel thermocouples are mounted between the cell and the shields to indicate the temperature differences between them. The adiabatic conditions of the cell are automatically controlled by two sets of temperature controller. The reliability of the calorimeter was verified through heat capacity measurements on the standard reference material α-Al 2 O 3 . The results agreed well with those of the National Bureau of Standards (NBS): within ±0.2% throughout the whole temperature region. The heat capacities of high-purity graphite and polystyrene were precisely measured in the interval 260–370 K by using the above-mentioned calorimeter. The results were tabulated and plotted and the thermal behavior of the two materials was discussed in detail. Polynomial expressions for calculation of the heat capacities of the two substances are presented.
What problem does this paper attempt to address?