[The Study of the MRI Artifacts Caused by Keepers].

Bin Wei,Fu-qiang Zhang,Qiang Yu
DOI: https://doi.org/10.3969/j.issn.1006-7248.2002.02.013
2002-01-01
Abstract:OBJECTIVE:The aim of this research was to observe the extension of MRI artifacts caused by keepers.METHODS:Five samples were imaged in a water phantom and a living human subject by 0.15 Tesla magnetic resonance devices. The maximum diameters of artifacts were measured and compared.RESULTS:The maximal diameters of the artifacts caused by disc-shaped and post-shaped keepers were 95.5mm and 139.4mm.CONCLUSION:Compared with the post-type keeper, the artifacts caused by disc-type keepers diminished significantly, and the difference of the maximal diameter reached statistical significance.