The Study of the MRI Artifacts Caused by Magnetic Attachments

WEI Bin,ZHANG Fu-qiang,YU Qiang
DOI: https://doi.org/10.3969/j.issn.1009-3761.2000.04.003
2000-01-01
Abstract:Objective: The aim of this research was to examine the difference of artifacts between whole magnetic attachment (magnetic assembly and keeper) and keeper upon MRI. Method: Three samples were imaged in a water phantom and a living human subject with 0.15 Tesla magnetic resonance devices. Results were analyzed and the maximum diameters of artifacts were measured. Results: The maximum diameter of the artifact caused by the whole magnetic attachment is 145.3mm, and that caused by keeper is 91.9mm. Conclusions: The scope of artifact caused by keeper is smaller than that caused by the whole magnetic attachment. The difference is statistically significant (p0.05).
What problem does this paper attempt to address?