Unfamiliar Artifacts and Elimination of MRI Installation and Adjustment

WANG Jin-xi,YUE Yun,WANG Peng-cheng,CHEN Fang,LIU Lin-xiang,WANG Wei-min
2008-01-01
Abstract:Objective: 5 examples of unfamiliar artifacts of 0.36T MRI at past 3 years presented,and the reason and solution were analysed.Result: The result shows that the main interference is from:a.External(not related to strict platform screen doors lead outside RF interference into the magnet space);b.indoor lights,laser positioning lights;RF and gradient coils(heat insulation material properties decreased after the instantaneous discharge);c.Cooling system components(especially water mains connector);signal acquisition of analog circuits(amplifiers,connect ware,filter plate);d.Digital Circuit(ADC,impon-derables less).Conclusion: We expect the work is benefit to magnetic resonance practitioners.