Research on 0.5T permanent magnet MRI system and MRI new technologies

Shanglian Bao,Jianying You,Qun He,Kaining Shi,Kun Zhou,Hongjie Zhang
DOI: https://doi.org/10.19650/j.cnki.cjsi.2010.03.023
2010-01-01
Abstract:A 0.5T permanent magnet MRI(Magnetic Resonance Imaging)system was developed,which includes the R & D of a human sized new structure magnet,a higher performance pre-amplifier/switch circuit,RF coils,temperature stability system,SE(Spin Echo),GE(Gradient Echo),FSE(Fast Spin Echo),perfusion,diffusion pulse sequences and new phase unwrapping & T1.T2 correction methodologies.We systematically studied the design and reconstruction method for parallel imaging technologies,completed the structure design for imaging guided intervention software system and realized some of these key technologies.The system specifications were measured and satisfied according to the national standard for QA(quality assurance)/QC(quality control).In commercial MRI,the pulse sequence of SSFP(Steady State Free Precession),new dynamical analysis methods for DCE-MRI(dynamic contrast-enhanced MRI),and spectroscopic limited DWI(diffusion-weighted imaging)were developed,which are on the way to be realized on 0.5T MRI system along with the progress of the magnetic field uniformity and system stability.All of these research results expand the investigation range and enhance the overall system to a higher level.
What problem does this paper attempt to address?