A Rapid Automated Shimming Method on Permanent Magnet MRI System

WANG Jin-xi,LIU Lin-xiang,LIU Tao,QIU Jian-feng,YUE Yun,WANG Wei-min
DOI: https://doi.org/10.3969/j.issn.1005-202X.2007.02.006
2007-01-01
Abstract:Objective: To develop a rapid and automated shimming method in permanent magnet MRI system.Methods: The inhomogeneity distribution of main magnet field B0 was achieved on two sets of image datas obtained by special imaging pulse sequence after the problem of phase rewind had been solved.The shimming current required to improve the field homogeneity was calculated and then loaded to the MR imaging system.The contrast experiment with and without the shimming current had been conducted on 0.36T permanent MRI system.Results:(1)The homogeneity of field in the view(FOV) was improved significantly.(2) MR signal and image quality were improved significantly.Conclusions: The shimming technology can adequately satisfy the clinical application.
What problem does this paper attempt to address?