Discussion of formation mechanism and solutions of artifacts commonly seen in magnetic resonance imaging examination

HE Weihong,FANG Xiangjun,CHEN Wei
DOI: https://doi.org/10.3969/j.issn.1673-7210.2011.35.045
2011-01-01
Abstract:Objective: To reduce the artifacts generated in magnetic resonance imaging(MRI) examination,improve the image quality and provide good image data for clinical use by studying the generation reasons and suppression methods of artifacts commonly seen in MRI examination.Methods: Medical records with different types and varying degrees of MR image artifacts of 168 patients of our hospital were collected for systematic analysis,of which 17 cases were limbs scan,37 cases were spine scan,71 cases were head and neck scan and 43 cases were abdomen scan.Results: There were 36 cases of folded artifacts,accounting for 2.4%,10 cases of volume effect artifacts,accounting for 5.9%,12 cases of chemical shift artifacts,accounting for 7.2%,66 cases of motion artifacts,accounting for 39.2%,24 cases of magnetic susceptibility artifacts,accounting for 4.3%,9 cases of cross artifacts,accounting for 5.4% and 11 cases of truncation artifacts,accounting for 6.6%.Conclusion: Understanding the generation mechanism of various artifacts and using various effective techniques to suppress or eliminate image artifacts are of important significance to improve the image quality and the diagnostic rate.