Evaluation of MRI artifacts

吴光耀,孙骏谟,黄雄
DOI: https://doi.org/10.3969/j.issn.1000-0313.2001.04.004
2001-01-01
Abstract:Objective:To understand the artifact′s characteristics and its causes and to investigate the methods to limit or eliminate the ghosts by analyzing common MRI artifacts.Methods:MR artifacts in 200 copies of films randomly selected,were analyzed and divided into four types.The followings were compared:artifact′s characteristics,the size of no signal region,geometric distortion,the degree of spatial blur and sequences with different parameters.Results:Motion artifacts(n=131)showed interleaved circular or arch ghost in the direction of phase.Susceptibility artifacts(n=25) including metal paramagnetic material artifacts (n=14) and poor homogeneity (n=11) caused by shimming and nonparamagnetic material,showed no signal or low signal area and figure distortion.Wraparound artifacts and partial volume effect (n=29) showed anatomic structure shift to next picture in the direction of phase, and the indistinct signal or shape of the lesion for partial volume effect.Among implement artifact and the miscellaneous (n=23),the most common artifact was chemical shift artifact,which showed black or white bandarch ghost along the surface between water and fat.Conclusion:The artifacts can be more effectively limited and eliminated by understanding the artifact′s appearances and its causes.
What problem does this paper attempt to address?