A Discrimination of Myocardial Tomography Artifacts Caused by Motion

石洪成,陈绍亮,李文罡,谷晓云,朱玮珉,姚之丰,刘文官
DOI: https://doi.org/10.3760/cma.j.issn.2095-2848.2001.02.006
2001-01-01
Abstract:Objective To detect the image characteristics of the motion artifacts, to determine the relationship among the axis, position and frames of the motion and the resultant image artifacts Methods A myocardial phantom was filled with the 99TcmO- 4 solution During imaging, the phantom was moved ±5 mm, ±10 mm, ±15 mm, ±20 mm, ±25 mm, ±30 mm and ±50 mm along the x、y and z axis respectively beginning at the different positions covering different ranges sequentially Every image was compared with normal image and assessed by three experienced observers without knowledge of the phantom motion The results were analyzed by logistic regression Results Motion artifact showed a radioactive nodule projecting from inferior wall while moved along x axis, symmetrical “hot areas showed on septal and lateral wall images while moved along y axis, and nodular hot areas showed in anterior wall while moved along z axis At last, all the motion artifacts became “double triangular The “lumpy and “defect areas showed oppositely in position while the motion covered the same distance but in the opposite directions along the axis Motion artifacts had something to do with the axis, distance and range of the frames affected, but had nothing to do with the beginning position of the motion Conclusions Motion artifacts had specific images when the phantom moved along different axes Frames of motion and motion along y axis contributed more to the motion artifacts than other factors