Correction of Cardiac Axial Movement During Myocardial Perfusion SPECT Imaging

XU Ying-jie,CAO Wen-tian,WANG Rong-fu,ZHANG Yan-yan,ZHANG Chun-li,ZHANG Wei-fang,LUO Jian-wei,CHEN Wen-bo
DOI: https://doi.org/10.13929/j.1003-3289.2012.12.050
2012-01-01
Abstract:Objective To correct the cardiac axial movement during myocardial perfusion SPECT imaging,in order to reduce the motion-artifact's influence on diagnosis.Methods Visual inspection of motion of the heart was done by means of cine of the raw data,sonogram and cyclogram.By cross-correlation,the camera angles were localized at which the motion occurred and determine the suggested correction for the motion.By means of visual inspection,the correction was modified and verified.Based in the correction at this stage,the images were reconstructed.The final motion correction was then determined by means of maximize the cross-correlation between the projection from the reconstructed image and the measured projection.Results Myocardial phantom showed that the motion of the phantom can be revealed at an accuracy of 3 mm for various kinds of axial motions.Motion artifacts on myocardial perfusion SPECT images were effectively improved.Conclusion This approach can be used to provide non-subjective,accurate,quantitative,and visual means of axial motion correction in myocardial perfusion SPECT imaging.