Study on best test samples of MRI artifacts caused by oral prostheses

WEI Bin,HU Yan-yan,YU Qiang,LUO Yi-xi,ZHANG Fu-qiang
DOI: https://doi.org/10.3969/j.issn.1674-8115.2008.11.001
2008-01-01
Abstract:Objective To investigate the best test samples of MRI artifacts caused by oral prostheses.Methods Three kinds of test samples were prepared with Ni-Cr alloy and 86.2%-Au alloy,and 3 raw material modules,cast modules and crowns were made,respectively.The samples were subjected to magnetic resonance imaging(MRI)examinations.The areas of MRI artifacts were measured,and SAS 8.2 statistic software was employed for analysis.Results The areas of MRI artifacts caused by Ni-Cr alloy raw material modules,cast modules and crowns were 543 to 17 177 pixels,110 to 5 050 pixels and 1 188 to 17 456 pixels,respectively,and those caused by 86.2%-Au raw material modules,cast modules and crowns were 396 to 5 267 pixels,151 to 1 271 pixels and 213 to 1 126 pixels,respectively.The areas of MRI artifacts caused by Ni-Cr alloy and 86.2%-Au raw material modules were significantly larger than those caused by cast modules(P<0.01).The areas of MRI artifacts caused by Ni-Cr crowns were significantly larger than those caused by cast modules(P<0.01).Conclusion There are significant differences in the areas of MRI artifacts caused by oral prostheses before and after cast.It is uncertain whether the appearance of samples could affect the areas of MRI artifacts.It is suggested that real prostheses in clinics should be chosen as samples for evaluation of MRI artifacts in order to assure the reliability of results.
What problem does this paper attempt to address?