The effects of the fixed prothesis made of Ni - Cr alloy on MRI

Yinglin SUN,Yuhua LIU
DOI: https://doi.org/10.3969/j.issn.1003-7632.2007.01.025
2007-01-01
Abstract:Objective To investigate the effect of the fixed prothesis made of Ni-Cr alloy on the MRI images.Methods Firstly, we made crowns with Ni-Cr alloy for a healthy volunteer's teeth. The crowns were divided into six groups, from group A to group F. Then they were put on teeth in order. Examined the volunteer's head and neck with a SIEMENS NOVUS 1.5T MRI scanner in axial and sagittal position. Standard SE (T_1) sequence was used to observe the the effects of each group. GRE sequence and EPI sequence were also used. The size of the artifacts were well recorded and analysed with eFilm medical software.Results As the amount of the crowns increases, the size of the distortion region increases, and the layers with artifacts also increase. The artifacts are limited in the lower maxillofacial area.Conclusion Fixed prothesis made of Ni-Cr alloy do create artifacts in MRI examination. However, the artifacts are limited in the lower maxillofacial area.