A Traceable Calibration Method for Spectroscopic Ellipsometry

Zhang Ji-Tao,Li Yan,Luo Zhi-Yong
DOI: https://doi.org/10.7498/aps.59.186
IF: 0.906
2010-01-01
Acta Physica Sinica
Abstract:A method for calibrating the spectroscopic ellipsometry by X-ray reflectivity is presented. As an indirect method,spectroscopic ellipsometry does not have the traceability because the measured film thickness is dependent on its optical constant. In contrast,at the grazing angle,the X-ray reflectivity can be used to measure the absolute thickness of thin film with sub-nanometer precision,and is independent of the optical constant. Five SiO 2 films were deposited on the substrate of single crystal silicon with the thickness of 2 nm,18 nm,34 nm,61 nm,and 170 nm,respectively. The results of spectroscopic ellipsometry and X-ray reflectivity were well fitted by a liner equation with slope of 1013±0013 and its intercept was set to zero,which means this calibration method is valid for spectroscopic ellipsometry in the determination of the thickness of thin films.
What problem does this paper attempt to address?