Calibrate the Non-Orthogonal Error of AFM with Two-Dimensional Self-Traceable Grating.

Wen Tan,Zhaohui Tang,Guangxu Xiao,Yulin Yao,Lihua Lei,Qing Li,Tao Jin,Xiao Deng,Xinbin Cheng,Tongbao Li
DOI: https://doi.org/10.1016/j.ultramic.2023.113734
IF: 2.994
2023-01-01
Ultramicroscopy
Abstract:The calibration of the non-orthogonal error in nanoscale measurements is of paramount importance for analytical measuring instruments. Particularly, the calibration of non-orthogonal errors in atomic force micro-scopy (AFM) is essential for the traceable measurements of novel materials and two-dimensional (2D) crystals. The 2D self-traceable grating with a theoretical non-orthogonal angle of less than 0.0027 degrees and an expanded uncertainty of 0.003 degrees (k = 2) are measured by the Metrological Large Range Scanning Probe Microscope (Met. LR-SPM). In this study, we characterized the local and overall non-orthogonal error in AFM scans and proposed a protocol to tune the optimal scanning parameters of AFM minimizing the non-orthogonal error. We presented the method for accurately calibrating a commercial AFM system for non-orthogonal by establishing a detailed un-certainty budget and errors analysis. Our results verified the important advantages of the 2D self-traceable grating in calibrating precision instruments.
What problem does this paper attempt to address?