Research on Comparison Measurement of Single-crystal Silicon Sphere Density Based on Pressure Floatation Method

刘子勇,王金涛,许常红,罗志勇,陈朝晖
DOI: https://doi.org/10.3969/j.issn.1000-1158.2013.02.01
2013-01-01
Abstract:To overcome the influence of liquid surface tension on the density measurement of single-crystal silicon sphere with hydrostatic method, pressure-of-flotation (PF) method and measurement system were developed to carry out the comparison measurement of single-crystal silicon sphere density. On the condition of certain temperature, the pressure of liquid was adjusted, which caused the change of liquid density, and the standard and measured single-crystal silicon spheres could be suspended in the liquid stably. By measuring liquid temperature, pressure and suspension heights, the density difference was determined. Double-level temperature control system was used to achieve stable temperature within ±0.25 mK. The liquid compress coefficient was determined by the linear equation of temperature-pressure suspension condition. It was verified by experiments that the density difference of single-crystal silicon spheres could be measured with high precision, and the standard relative uncertainty is 2×10-7.
What problem does this paper attempt to address?