Development and Verification of High Accuracy Apparatus for Solubility Measurement

Xiaolong Cui,Guangming Chen,Qin Wang,Changsheng Li,Xiaohong Han
DOI: https://doi.org/10.3321/j.issn:0253-987X.2007.07.024
2007-01-01
Abstract:A new apparatus was developed to investigate solubility characteristics of refrigerant-absorbent. It consists of a temperature measurement system (253.15-363.15 K, temperature uncertainty less than ±14 mK), a pressure measurement system (0-3.5 MPa, pressure uncertainty less than ±1.6 kPa), a water thermostat bath (293.15-363.15 K), an alcohol thermo stat bath (253.15-293.15 K), a vacuum system, and a mixture make-up system. In order to check the reliability and accuracy of the system, several test measurements were conducted on the vapor pressure of R134a and p-T diagram for the binary mixture of R22+DMF. The test results are in good agreement with the reference data.