An Apparatus for Determination of Mutual Solubilities in Near-critical Water

吕秀阳,郑赞胜,何龙,鹿骋,夏文莉
DOI: https://doi.org/10.3321/j.issn:1003-9015.2004.05.001
2004-01-01
Abstract:Lack of thermodynamic data is one of the factors hindering the research and development of green chemical processes in near-critical water. An apparatus was constructed for determination of mutual solubilities of water-organics in near-critical water. It can be used at temperature up to 350°C and pressure up to 30 MPa. The apparatus makes the features of small pressure variation, precise temperature control, on-line sampling and resistance to corrosion. The reliability of the apparatus was examined by the comparison of experimental data with literature ones and by the reproducibility of experimental data. The apparatus was used to systematically determine the mutual solubilities of water-benzene system in the temperature range of 100-300°C and pressure up to 30 MPa.
What problem does this paper attempt to address?